This content
provides you with a example Radiographic
Testing Procedure. This is a general and sample RT procedure and you need to
modify it to meet your project specifications.
SCOPE:
This
Procedure describes the general requirements for radiography examination (RT)
according to related approved weld map for the metallic welding and casting as
may be required by the specification or under which component is being designed
and manufactured.
This
radiographic testing procedure provides the material, equipment, calibration,
personnel qualification, examination process, evaluation, records and
acceptance standards forXXXProject which will be
fabricated in YYY.
SURFACE CONDITION
According to
T.222.2, the weld ripples or weld surface irregularities on the both the inside
(where accessible) and out side shall be
removed by any suitable process to such a degree that the resulting radiographic
testing image due to any surface irregularities cannot mask or be confused with
the image of any discontinuity. the finished surface of all butt welded joints
mat be flush with the base material or may have reasonably uniform crowns, with
reinforcement not to exceed that specified in the referencing code section.
RADIATION SOURCE
X-Radiation:
The
radiography testing techniques shall demonstrate that the required radiography
sensitivity has been obtained. Maximum x-ray voltage is 300 KV.
Gamma radiation:
The
recommended minimum thickness for which Radio-active isotopes may be used as
follow:
Table – 3.2
Material |
Iridium192 |
Cobalt 60 |
Steel |
0.75 in |
1.50 in |
Copper or high nickel copper |
0.65 in |
1.30 in |
Aluminum |
2.50 in |
-- |
The maximum
thickness for the use of radioactive isotopes is primarily dictated by exposure
time, therefore; upper limits are not shown. The minimum Recommended thickness
limitation may be reduced when the radiography techniques are used to
demonstrate that the required radiographic testing sensitivity have been
obtained, by purchaser approval.
RADIGRAPHIC
FILMS
Any
commercially available industrial radiography films may be used in accordance
with SE 1815(ASTM) standard test method for film system in industrial
radiography. Radiography film shall be fine grain high definition, high
contrast film (Kodak type AA 400, FUJI 100 or AGFA D7).
SCREENS
Any
commercially available intensifying screen, except those of the fluorescent
type, may be used. Intensifying screen for x-ray or Gama ray method divided in
two categories: 1-front screen 2-back screen. Commonly lead screens use with 27
micron thickness. (Front screen)
PENETRAMETER (I.Q.I)
Penetrameters shall be either the whole type or the wire type and
shall be manufactured and identified in accordance with the requirements or
alternatives allowed in SE 142 or SE 1025 (for whole type) and SE-747 (for wire
type), and appending. ASME V 2007 ED & ASME Sec VIII Div I ED 2007.
Penetrameters shall consist of those in table 233.1 for hole type and those in table 233.2 for wire type. (Wire
type IQI shall be used for welds.)
SELECTION OF PENETRAMETER (I.Q.I)
Material.
IQIs
shall be selected from either the same alloy material group or grade as
identified inSE-1025, or SE-747, as applicable, or from an alloy material Group
or grade with less radiation absorption than the material being radiographed.
Size.
The
designated hole IQI or essential wire listed
in Table T-276 provided an equivalent IQI sensitivity is maintained. See
T-283.2.shall be as specified in Table T-276. A thinner or thicker hole-type
IQI may be substituted for any section thickness
(a) Welds
With Reinforcements. The thickness on which the IQI is based is the nominal
single-wall thickness plus the estimated weld reinforcement not to exceed the
maximum permitted by the referencing Code Section. Backing rings or strips
shall not be considered as part of the thickness in IQI selection. The actual
measurement of the weld reinforcement is not required.
(b) Welds
Without Reinforcements. The thickness on which the IQI is based is the nominal
single-wall thickness. Backing rings or strips shall not be considered as part
of the weld thickness in IQI selection.
Welds Joining Dissimilar Materials or Welds with Dissimilar Filler Metal.
When the weld
metal is of an alloy group or grade that has a radiation attenuation that
differs from the base material, the IQI material selection shall be based on
the weld metal and be in accordance with T-276.1. When the density limits of
T-282.2 cannot be met with one IQI and the exceptional density area is at the
interface of the weld metal and the base metal, the material selection for the
additional IQIs shall be based on the base material and is in accordance with
T-276.1
PLACEMENT OF RADIOGRAPHIC TESTING PENETRAMETER (I.Q.I)
Source side penetrameters:
The penetrameters shall be placed on the source side of
the part being examined, except for the condition described in chapter 8.2.
film side penetrameters:
Sensitivity:
The
sensitivity required using wire type IQI shall be 2%.
Sensitivity:(Diameter of thinnest wire visible on radiograph / Part
thickness at IQI location) x
100
Where
inaccessibility prevents hand placing the penetrameter (s)
on the source side, it shall be placed on the film side in contact with the
part being examined. A lead letter “F” shall be placed adjacent to or on
the penetrameter (s).
NUMBER OF PENETRAMETER (I.Q.I)
When one or
more film holders are used for an exposure, ate least one penetrameter imager shall appear on each radiograph.
If the
requirements of T-282 are met by using more than onepenetrameter,
one shall be representative of the lightest area of interest and the other the
darkest area of interest.
The
intervening densities, on the radiograph, shall be considered as having
acceptable density.
Number of
I.Q.I shall be according to ASME SEC V.T.277.2.
RADIOGRAPHIC TESTING TECHNIQUE
A single-wall
exposure technique shall be used for radiography whenever practical. When it is
not practical to use a single-wall radiographic testing technique, a
double-wall technique shall be used. An adequate number of exposures shall be
made to demonstrate that the required coverage has been obtained.
Single-Wall Technique.
In the
single-wall radiographic testing technique, the radiation passes through only
one wall of the weld (material), which is viewed for acceptance on the
radiograph.
Double-Wall Technique.
When it is
not practical to use a single-wall technique, one of the following double-wall
techniques shall be used.
(a)
Single-Wall Viewing. For materials and for welds in components, a technique may
be used in which the radiation passes through two walls and only the weld
(material) on the film-side wall is viewed for acceptance on the radiograph.
When complete coverage is required for circumferential welds (materials), a
minimum of three exposures taken 120 deg to
each other shall be made.
(b)
Double-Wall Viewing. For materials and for welds in components 31⁄2 in.
(89 mm) or less in nominal outside diameter, a technique may be used in which
the radiation passes through two walls and the weld (material) in both walls is
viewed for acceptance on the same radiograph .For double-wall viewing, only a
source-side IQI shall be used. Care should be exercised to ensure that the required
geometricunsharpness is not exceeded. If the
geometric unsharpnessrequirement cannot be met,
then single-wall viewing shall be used.
(1) For
welds, the radiation beam may be offset from the plane of the weld at an angle
sufficient to separate the images of the source-side and film-side portions of
the weld so that there is no overlap of the areas to be interpreted. When
complete coverage is required, a minimum of two exposures taken 90 deg to each other shall be made for each joint.
(2) As an
alternative, the weld may be radio graphed with the radiation beam positioned
so that the images of both walls are superimposed. When complete coverage is
required, a minimum of three exposures taken at either 60 deg or 120 deg to
each other shall be made for each joint.
SOURCE TO OBJECT AND OBJECT TO FILM DISTANCE (SOD & OFD)
According to
geometric unsharpness formula (Ug = f *OFD/FOD) for minimizing the Ug value, OFD value shall be minimizing therefore
object to film distance shall be minimum.
Source to
object distance (SOD) shall be set according radiographic technique, object
shape and strength of source.
RADIOGRAPHIC TESTING IDENTIFICATION SYSTEM
The method
shall be used to produce permanent identification to the radiographies
traceable to the contract, components, welds or weld seams, or part numbers, as
appropriate. This identification mark shall not obscure the area of interest.
RADIOGRAPHIC TESTING ACCEPTANCE STANDARD
Refer to ASME
Sec VIII, Div. I
a) Butt
welded joints surfaces shall be sufficiently free from coarse ripples ,grooves
, overlaps and abrupt ridges and valleys to permit proper interpretation of
radiographic and the required non-destructive examinations.
If there is a
question regarding the surface condition of the weld when interpreting a
radiographic film, the film shall be compared to the actual weld surface for
determination of acceptability.
b)
Indications shown on the radiographies of welds and characterized as
imperfections are un-acceptable under the following condition:
1) Any
indications characterized as a crack or zone of incomplete fusion or
penetration.
2) Any other
elongated indication at radiography, which has length greater than:
(a) 1/4 in.
(6mm) for t up to 3/4 in. (19mm)
(b) 1/3 t for
t from 3/4 in. (19mm) to 2 1/4 in. (57mm)
(c) 3/4 t
(19mm) for t over 2 1/4 in. (57mm)
Where: t=
thickness of weld excluding any allowable reinforcement.
3) Any group
of aligned indications that have an aggregate length greater than t in a length
of 12t , except when the distance between
the successive imperfections exceed 6L where L is the length of the longest
imperfection in the group.
4) Rounded
indications in excess of that specified by the acceptance standards given in
ASME sec. VIII, DIV I, appendix 4 fig. 4-2 to 4-8
Note: spot RT
shall be done as per ASME Sec. VIII, Div. 1 UW-52; however the acceptance
criteria shall be according to UW-51 (as
specification).
DEFECT REMOVAL
Repair area
shall be located on the weld line after evaluation & interpretation of
radiograph .defects shall be removed by suitable method such as grinding,
chipping or gouging (if permitted).welding of the repair area shall meet the
requirement of related WPS,PQR.
CERTIFICATION AND PERSONNEL QUALIFICATION IN RADIOGRAPHIC TESTING.
Personnel
performing radiography examination to this procedure shall be qualified and
certified by XXX also shall meet the requirements of ASNT-SNT-TC-1A-2001
EDITION at least level II and on ASNT-SNT-TC-IA for code section I and sec VII
div 2. Film interpreter shall have level II as a minimum.