What is a Limitation of Using
Electron Microscopes to View Specimens?
How does electron microscopy work?
Three
main factors underlie the scientific principle of microscopy: magnification,
illumination, and resolution.
- Resolution is
perhaps the central tenet of microscopy in that it defines the ability of
any device to distinguish between two closely spaced objects. In other
words, the ability to reveal structural details as distinct and separate.
- Diffraction (or
the mathematically defined Diffraction Limit)
is used as the measure of resolution and is impacted by factors including:
the wavelength of incident light, the refractive index of materials
through which the light passes, and the numerical aperture of the lens
used to magnify the image.
·
Consequently,
there is a limit (the Diffraction Limit) beyond which the ability to resolve
two images becomes impossible. The lowest diffraction achievable with
convention lenses is 200 nm, and efforts to surpass this point have been
limited even with modern techniques, although there are notable exceptions.
How do SEM
electron microscopes work?
The
electron microscope concept overcomes this diffraction limit and permits
imaging of much smaller and complex substances and biological specimens.
Scanning Electron Microscopy (SEM)
SEM
involves the production of an image by scanning the specimen with a focused
beam of electrons.
- These electrons interact with electrons in
the sample and produce signals representative of the sample topography and
composition.
- The beam position is closely linked with
the emission signal to construct a resulting image in fine detail – at a
resolution in the low nm range.
- The most commonly used method in biological
applications is secondary electron detection from excited atoms within the
beam.
- Others include back-scattered or reflected
electrons and cathodoluminescence. It is rare, however, that a SEM
instrument can detect more than one signal type simultaneously.
The
image is produced from electronic interactions at various depths of the
specimen, and secondary electron imaging (SEI), the most common method, is
performed by detection at the very surface.
- The result is vey high 1 nm surface
resolution -- although subsurface detail is precluded.
- To this end, back-scattered electrons
(BSE), involving interactions deeper within the sample and through a
separate electronic mechanism, can provide subsurface identification of
the elemental composition of a sample.
Other traits of SEM include a
large depth of field, yielding a 3-dimensional appearance, and a wide range of
magnifications – 10x to 500,000x, or approximately 250 times that of the best
light microscopes.
What is a
limitation of using SEM electron microscopes to view specimens?
The
power of SEM does come with certain requirements for sample preparation.
- Specimens must be both electrically
conductive at the surface and electrically grounded to prevent
accumulation of static charge.
- Nonconductive samples can be coated with a
conductive film such as gold or palladium.
- Alternatively, environmental or low-voltage
can be used on native non-conducting samples, although these may require
specific (and sometimes tedious) conditions.
- Biological samples have a requirement to be
completely dry due to high vacuum conditions of the mounting chamber.
Therefore, most samples involving live tissues and cells must be
chemically fixed by treatment with glutaraldehyde or formaldehyde to
preserve native structure.
- SEM samples can also be cold treated in a
process called cryofixation (typically involving a SEM stage built for
this) and can be imaged by low-temperature scanning.
- Samples can also be freeze-fractured to
reveal inner structures in fine detail.
The
resolution in SEM is not limited by the diffraction limit or the quality of
lenses or mirrors, etc.
- Rather, spatial resolution in SEM depends
on the size of the electron beam and the volume of specimen material
interacting with the beam -- or interaction volume.
- As a result, the resolution falls between 1
nm and 20 nm. Impressive indeed, although not as high resolution of the
next technique to be discussed, which can measure distances between
individual atoms in a sample.
How do TEM
electron microscopes work?
Transmission Electron Microscopy (TEM)
TEM
is similar to SEM in that a narrow beam of electrons is projected onto a
sample.
- A major difference, however, is that the
electronic interactions are captured after transmission through the thinly
sliced sample, after which the signal is amplified and then interpreted to
yield structural information.
- TEM is capable of imaging at the single
electron level, many thousands of times smaller that light microscope
limits.
- Various imaging modulations can be
performed to account for the thickness and condition of the sample.
There are, however, severe
limitations to the types of material that can be imaged and the conditions
required. Regardless, TEM is a powerful technique that has yielded significant
results in fields as diverse as virology, materials science, and
nanotechnology.
What is a
limitation of using TEM electron microscopes to view specimens?
With
regard to TEM:
- Preparation of cells for TEM requires fixation
(with chemicals or cryotechniques) that can introduce artifacts and
damage.
- The preparation of cells/tissues for TEM is
lengthy.
- TEM relies on highly sophisticated and
expensive machinery.
Regarding Cryo-TEM:
- The samples are unstained, therefore, the technique
produces low-contrast images.
- Thin sectioning of samples in a frozen
state is a technical and tedious procedure.
- Cryo-TEM relies on highly sophisticated and
expensive machinery.
Two
considerations dominate the discussion:
- Although a powerful, high resolution
technique, traditional and cryo-TEM require extensive and tedious sample
preparation.
- This often involves ultra-thin sectioning
by use is a microtome and sample fixation, both of which can introduce
artifacts and can damage specimens.
- The second consideration is the large investment in
obtaining and maintaining a TEM instrument.